Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Following information is part of the intended use according to /wiki/spaces/QA/pages/1800011780 , Annex II, 1.1 c and , /wiki/spaces/QA/pages/176685192 , 5.2.

Intended Medical Indication / Claim

The product is used to present information on the performance of the peripheral vestibular function by providing objective measures of eye-movement with or without stimulation of the balance organ in general as well as eye-velocity response to head-velocity stimulus, showing the VOR gain in the plane of rotation of the head.

It enables the recording of eye movements in response to head movements without making a diagnostic statement about the user's state of health. The results are intended to serve as a tool for professionals to observe and analyze the functionality of the vestibular system.

Contraindication

  • Blindness

  • Broken nose or other face/head trauma

  • Problems with the cervical spine, e.g., neck trauma, acute cervical disc herniation

  • Recent eye surgery

  • Ptosis on measurement eye

  • Excessive eye blinks

  • Excessive eye make up

  • Contact with insured skin

Patient Population

The product is intended for patient from 6 years and up. The patient must be physically sound as the procedures include some physical stimulation/movement. Patient must be able the see the target without the use of spectacles.