Zeisberg GmbH

Skip to end of metadata
Go to start of metadata

You are viewing an old version of this content. View the current version.

Compare with Current View Version History

« Previous Version 3 Next »

Following information is part of the intended use according to /wiki/spaces/QA/pages/1800011780 , Annex II, 1.1 c and , /wiki/spaces/QA/pages/176685192 , 5.2.

Intended Medical Indication / Claim

The product is used to present information on the performance of the peripheral vestibular function by providing objective measures of eye-movement with or without stimulation of the balance organ in general as well as eye-velocity response to head-velocity stimulus, showing the VOR gain in the plane of rotation of the head.

It enables the recording of eye movements in response to head movements without making a diagnostic statement about the user's state of health. The results are intended to serve as a tool for professionals to observe and analyze the functionality of the vestibular system.

Contraindication

  • Blindness

  • Broken nose or other face/head trauma

  • Problems with the cervical spine, e.g., neck trauma, acute cervical disc herniation

  • Recent eye surgery

  • Ptosis on measurement eye

  • Excessive eye blinks

  • Excessive eye make up

  • Contact with insured skin

Patient Population

The product is intended for patient from 6 years and up. The patient must be physically sound as the procedures include some physical stimulation/movement. Patient must be able the see the target without the use of spectacles.