Setting up the KA450 appliance

Zeisberg GmbH

Setting up the KA450 appliance

Please note!

Do not place the appliance in one place:

  • where it is exposed to direct sunlight

  • where it is exposed to moisture

  • where it is placed close to heat sources such as radiators

  • which is poorly ventilated

There are different ways of setting up the KA450 depending on the space available:

  • The patient and the examiner sit directly opposite each other or offset, with the examiner having the device directly in front of them

  • The patient sits in a hearing test booth. The examiner sits outside the booth and has the audiometer directly in front of them. The examiner has eye contact with the patient through a window in the booth.

Please note!

In order to achieve reliable audiometric measurement results on test subjects, the KA450 must be set up in a room that is as free as possible from background noise. The best measurement results are obtained by using a hearing test booth. Ask your sales partner.

Setting up the speakers

The free-field loudspeaker boxes must be directed towards the patient at an angle of 90° to the sound emission surface and must be positioned at head height to the patient. The distance between the loudspeaker box and the patient should be in the range of 1 - 1.5 m.

Related content

Installation of KA450 free-field audiometry
Installation of KA450 free-field audiometry
More like this
Setting up the KA450 speakers
Setting up the KA450 speakers
More like this
Introduction KA450
Introduction KA450
More like this
Sound audiometry, air line KA450
Sound audiometry, air line KA450
More like this
Connecting the audiometer to a KA450 soundproof booth
Connecting the audiometer to a KA450 soundproof booth
More like this
Features of the KA450 sound field
Features of the KA450 sound field
More like this