Safety check (STK) KA450
Zeisberg GmbH
Safety check (STK) KA450
In order to ensure the safety of the KA450 over its service life, we recommend that an additional safety check is carried out annually as part of the measurement check. The safety check includes the following tests:
Checking the protective conductor resistance
Measurement of the insulation resistance
Measurement of the earth leakage current
Measurement of the housing leakage current
Measurement of the patient leakage current
Measuring the power consumption of the device
Test report on the measurements carried out for filing in the device logbook