Anesthesia during measurement of bone conduction KA450

Zeisberg GmbH

Anesthesia during measurement of bone conduction KA450

The crosstalk attenuation from one ear to the other is a maximum of 5 to 10 dB when measuring with the bone conduction receiver, and it is therefore assumed for safety's sake that the signal can also be heard by the opposite ear without attenuation.

Therefore, it is always necessary to anesthetize the opposite ear during bone conduction measurements, and it must be performed at least at the volume of the test signal if the opposite ear is completely intact.

Deafening is performed if the bone conduction on one side deviates by 10 dB or more, or if there is a bone conduction-air conduction difference of 15 dB or more.

First, the bone receiver is placed on the mastoid. The air receiver is used for anesthesia as follows:

One air receiver capsule is placed on the ear to be deafened. The other air receiver capsule is placed on the side of the test ear in the temple area so that the auditory canal remains open.

 

Note:             

In the case of bone conduction, the auditory canal of the test ear must not be covered, otherwise the bone conduction will be perceived too well and incorrect measurement results will be obtained (à Weber effect).

 

As with the air duct measurement, the masking signal is usually a narrow band noise (NBN = Narrow Band Noise), the volume of which can be adjusted using the level control (42, 47).

Depending on the hearing loss of the ear to be masked, a correspondingly louder masking sound should be selected compared to the test signal.

The following values can be used as a rule of thumb for anaesthetization with bone conduction:

70 dB for minor hearing loss

80 dB for moderate hearing loss

90 dB for severe hearing loss

Related content

Tone audiometry, bone conduction KA450
Tone audiometry, bone conduction KA450
More like this
Sound audiometry, air line KA450
Sound audiometry, air line KA450
More like this
Anschluss des Audiometers an eine Schallschutzkabine CA350/CA450
Anschluss des Audiometers an eine Schallschutzkabine CA350/CA450
More like this
Anschluss des Audiometers an eine Schallschutzkabine KA450
Anschluss des Audiometers an eine Schallschutzkabine KA450
More like this
Atomization when measuring the air duct KA450
Atomization when measuring the air duct KA450
More like this

Gutenbergstrasse 39 * 72555 Metzingen

Achtung: Ausgedruckte gelenkte Dokumente sind nicht gültig!
Die gültige Fassung ist in Confluence abrufbar.