Zeisberg GmbH
Connecting an external CD player
First connect the signal cable to the CD player. Then connect the white or black Connection plug in the rear connection socket "Signal input left" (12) is plugged in. The red The connector plug is inserted into the rear connection socket "Signal input right" (11) is plugged in.
For the connection of the power supply, other electrical connections and safety instructions for the CD player, please refer to the manufacturer's accompanying documents.
Please note!
After installing the CD player or after each CD change, check the VU meter settings for the signal inputs in the Setup menu Settings > VU meter with the CCITT noise of the CD used.
Warning!
When using an external CD player or tape recorder, ensure that they comply with the provisions of EN 60601-1. If necessary, they must be operated via an isolating transformer or power supply unit in accordance with EN 60601-1 (see also the chapter "Combination with other devices").