Noise level KA450

Zeisberg GmbH

Noise level KA450

The maximum permissible noise level in the test room varies depending on the required measuring ranges:

  • If frequencies down to 125 Hz are to be measured, the permissible noise levels in the low-frequency range are somewhat lower than if you can restrict yourself to frequencies >= 250 Hz.

  • If the hearing threshold is to be determined down to 0 dB, the permissible noise levels are lower than if the hearing threshold is determined down to +5 dB.

 

Third-octave band center frequency

Noise level limit / [dBSPL]

Hearing threshold determination up to 0 dB

Hearing threshold determination up to +5 dB

(à +8 dB compared to 0dB hearing threshold)

Lowest frequency of the test signal 125 Hz

Lowest frequency of the test signal 250 Hz

Lowest frequency of the test signal 125 Hz

Lowest frequency of the test signal 250 Hz

31,5

52

60

60

68

40

44

53

52

61

50

38

46

46

52

63

32

41

40

49

80

27

36

35

42

100

22

32

30

40

125

17

25

25

33

160

14

18

22

26

200

12

12

20

20

250

10

10

18

18

315

8

8

16

16

400

6

6

14

14

500

5

5

13

13

630

5

5

13

13

800

4

4

12

12

1000

4

4

12

12

1250

4

4

12

12

1600

5

5

13

13

2000

5

5

13

13

2500

3

3

11

11

3150

1

1

9

9

4000

-1

-1

7

7

5000

1

1

9

9

6300

6

6

14

14

8000

12

12

20

20

10000

14

14

22

22

12500

15

15

23

23

Related content

Sound audiometry KA450
Sound audiometry KA450
More like this
Sound audiometry, air line KA450
Sound audiometry, air line KA450
More like this
Setting up the KA450 appliance
Setting up the KA450 appliance
More like this
Technical data KA450
Technical data KA450
More like this
Technische Daten KA450
Technische Daten KA450
More like this
Features of the KA450 sound field
Features of the KA450 sound field
More like this

Gutenbergstrasse 39 * 72555 Metzingen

Achtung: Ausgedruckte gelenkte Dokumente sind nicht gültig!
Die gültige Fassung ist in Confluence abrufbar.