Metrological control (MTK) KA450
The device is subject to the Medical Devices Operator Ordinance (MPBetreibV). According to §11 (MPBetreibV), the KA450 must therefore be subjected to an annual metrological check. The metrological check includes the following tests:
Function check
Visual inspection
Check and, if necessary, adjust the pressure force of the headband
Testing the frequency accuracy of the test tones
Distortion factor measurement of air and bone conductors
Checking and, if necessary, adjusting the test levels for air and bone conduction at all test frequencies
Attenuator test
Measurement of free field equalization (if required)
Checking and, if necessary, adjusting the test levels for the air duct and free field for speech
Final subjective audiometer check
Test report on the measurements carried out for filing in the device logbook